Quantitative Composition Analysis of Modified Plastics Using Thermogravimetric Analyzer (TGA)
Quantitative Composition Analysis of Modified Plastics Using Thermogravimetric Analyzer (TGA)
Modified plastics are widely used in modern industry, and their performance and quality are closely related to their composition. Accurate determination of the composition of modified plastics is of great significance for evaluating performance, optimizing production processes, and quality control. As an important thermal analysis technique, the thermogravimetric analyzer (TGA), with its ability to accurately monitor mass changes of materials during heating, provides key information for quantitative anal...
I. Testing Principle
The basic principle is to place the sample on a high-precision balance and expose it to a controlled atmosphere and programmed temperature. As the temperature rises, the sample undergoes physical or chemical changes, resulting in an increase or decrease in mass. By analyzing the relationship between mass changes and temperature or time, detailed information about the thermal behavior and reaction mechanisms of the sample can be obtained.
II. Experimental Steps
1. Sample Preparation
1.1 Sample form treatment: Depending on the initial form of the sample, different treatments are applied. Powder materials should be evenly spread into a thin layer at the bottom of the crucible. To minimize particle size effects, sample particles should be as small as possible. Bulk materials should be cut into thin slices or crushed into granules and evenly distributed at the crucible bottom. Fibrous materials can be cut into small segments, evenly dispersed, or wound around a small rod, then placed in...
1.2 Sample mass control: Typically controlled between 10–20 mg. The sample should be evenly distributed, avoiding agglomeration, filling about one-third to one-quarter of the crucible volume. Moisture-absorbing or volatile materials should undergo pretreatment (e.g., drying).
2. Testing Instrument
- Instrument name: TGA1000 Thermogravimetric Analyzer
- Instrument brand: Nanjing Dazhan Instrument Testing Co., Ltd.
2.3 Instrument calibration:
Before testing, the thermogravimetric analyzer must be calibrated to ensure accuracy. Standard reference materials are used to calibrate by comparing their known thermogravimetric curves with measured results, adjusting instrument parameters accordingly.
3. Experimental Parameter Settings
- Temperature program: According to the characteristics and test purpose of modified plastics, appropriate heating rates (commonly 10–20 ℃/min) and temperature ranges are set. Generally, the temperature is raised to 800 ℃ with a heating rate while passing gas (e.g., nitrogen or air) at about 50 mL/min.
- Atmosphere selection: The heating process is carried out entirely under nitrogen protection.
4. Experimental Operation
Place the prepared high-temperature engineering plastic sample crucible into the sample chamber of the thermogravimetric analyzer, ensuring stable placement. Start the instrument, and according to the programmed heating schedule, it continuously records the mass change of the sample, generating thermogravimetric (TG) and derivative thermogravimetric (DTG) curves. During the experiment, the instrument status and curve changes should be closely monitored to ensure smooth operation.
III. Data Analysis
By observing the thermogravimetric curve, the mass change of the sample at different temperature intervals is determined. Given the multi-component characteristics of modified plastics, a complex program with multiple heating and isothermal stages is often designed, with atmosphere switching at different stages.
A typical test procedure includes three core stages:
- Dehydration/devolatilization stage: Room temperature to ~150 ℃ under N₂ atmosphere, removing adsorbed water and volatile substances.
- Polymer pyrolysis stage: 150–600 ℃ under inert atmosphere, allowing complete decomposition of the resin matrix.
- Filler decomposition/oxidation stage: 600–800 ℃ under oxidative atmosphere, where carbon black or inorganic fillers are oxidized and decomposed.
IV. Experimental Conclusion
The thermogravimetric analyzer has important application value in the quantitative analysis of modified plastic compositions. By accurately measuring mass changes during heating and applying reasonable data analysis methods, quantitative determination of the components in modified plastics can be achieved. This provides strong technical support for research, production, and quality control of modified plastics. With the continuous development and improvement of thermogravimetric analysis technology, as w...